ozone bagging
【产品名称】臭氧套袋
【型号】 YH-200-A、YH-200-B、YH-300-A、YH-300-B、YH-300H-A、YH-300H-B、YH-400-A、YH-400-B、YH-400-C、YH-400-D、YH-400-E、YH-400-F、YH-400H-A、YH-400H-B、YH-400H-C、YH-400H-D。
The specific product model description is shown in the table below:
|
Model |
Application site |
Bag length × width/ (mm) |
Model |
Application site |
Bag length × width/ (mm) |
|
YH-200-A |
head |
350×290 |
YH-400-C |
calf |
335×650 |
|
YH-200-B |
head |
360×310 |
YH-400-D |
calf |
345×680 |
|
YH-300-A |
forearm |
200×590 |
YH-400-E |
foot |
335×320 |
|
YH-300-B |
forearm |
210×620 |
YH-400-F |
foot |
345×330 |
|
YH-300H-A |
forearm |
245×570 |
YH-400H-A |
calf |
425×670 |
|
YH-300H-B |
forearm |
255×590 |
YH-400H-B |
calf |
445×700 |
|
YH-400-A |
calf |
335×650 |
YH-400H-C |
foot |
370×315 |
|
YH-400-B |
calf |
345×680 |
YH-400H-D |
foot |
380×330 |
[Product Performance] When inspected with normal or corrected vision, the bag body should have no obvious burrs or stains, and the welding lines on the bag body should be tight, clear, and uniform, without any gaps or other defects. The surface of the tubing should be smooth and free of kinks; the bag should be able to withstand a pressure of 2KPa without leakage; when the liquid stop clamp is in the closed state, it should be able to close the tubing passage.
[Main structural components] It consists of a sealing ring, a bag body, an air circuit interface, and a pipeline. It is used in conjunction with a medical ozone generator or medical ozone therapy device and is placed over the patient's treatment area. It is not a sterile product.
[Scope of Application] Used to assist in the introduction of medical ozone.
[Contraindications] For individuals allergic to ozone or those with asthma, this product is contraindicated.
[Contents of precautions, warnings, and tips]
Note: Please follow medical advice throughout use. Any potential impact from improper use is unrelated to the product;
(1) This product serves only as an auxiliary device and does not replace the core functions of ozone therapy equipment;
(2) Please keep this product out of reach of children, and children should use it under the supervision of a guardian or medical personnel;
(3) This product should be disposed of in accordance with the Regulations on the Management of Medical Waste after use.
[Installation and Usage Instructions]
(1) Before use, check whether the product's outer packaging is damaged and whether it is within its expiration date;
(2) Pull open the sealing ring with both hands and slowly insert the affected side into the bag, being careful not to touch the wound;
(3) One end of the pipeline is connected to the gas circuit interface, and the other end is connected to a medical ozone generator or medical ozone therapy equipment;
(4) Continuously inject ozone into the bag;
(5) After inflation, remove the tubing and close the gas circuit connector.
[Storage and Transportation Conditions] This product should be stored in a clean, dry, ventilated, and shaded place, away from heat sources, and isolated from toxic, harmful, and corrosive substances. [Shelf Life] Under the conditions of storage and transportation, the product has a shelf life of 3 years
[Medical Device Registration Certificate Number] JI Lang Xiang Bei 20260003